MC-3001 Split-Type Coating Thickness Gauge
Other Products
Product Overview
The MC-3001 split-type external probe coating thickness gauge can quickly, non-destructively, and precisely measure the thickness of coatings and platings. It is suitable for both laboratory and on-site engineering applications. By using different probes, it can meet various measurement needs. This instrument is widely used in manufacturing, metal processing, chemical industry, commodity inspection, and other testing fields. It is an essential instrument for material protection professionals. This instrument complies with the following standards: GB/T 4956─2003 Magnetic method for measuring the thickness of non-magnetic coatings on magnetic substrates; GB/T 4957─2003 Eddy current method for measuring the thickness of non-conductive coatings on non-magnetic metal substrates.
II. Features
● Utilizes magnetic thickness measurement method to non-destructively measure the thickness of non-magnetic coatings (e.g., zinc, aluminum, chromium, copper, rubber, paint) on magnetic metal substrates (e.g., steel, iron, alloy, hard magnetic steel); employs eddy current thickness measurement method to non-destructively measure the thickness of non-conductive coatings (e.g., rubber, paint, plastic, anodized film) on non-magnetic metal substrates (e.g., copper, aluminum, zinc, tin).
● One host can be freely paired with probes and supports up to 6 types of probes.
● External split probe with stable body design.
● High-definition color screen for clearer display.
● Measurement modes: Simplified mode, monitoring mode, statistical mode, minimum value capture.
● Temperature compensation: Equipped with the latest "phase-locked loop" technology.
● Screen flip: Manual selection for flipping and displaying measurement data.
● Display method: 2.4-inch TFT high-definition color screen, 320*240 pixels, large digital display.
● Measurement speed: Single and continuous measurement speeds.
● Calibration methods: Substrate calibration, system calibration, one-point calibration.
● Limit setting: Automatically alerts for measurements outside the set limits.
● Display language: Built-in Chinese and English.
● Measurement units: Metric and imperial convertible.
● Low battery indicator function.
● Audible beep prompts during operation.
● Storage function: Manual and automatic storage options, with capacity for over 10,000 measurement points.
● Delete function: Delete a suspicious data point from a stored group or all stored data for new measurements.
● Wireless communication function: Bluetooth 2.0 for wireless communication with computers or printers.
● Two shutdown methods: Manual and automatic shutdown.
● V-groove: Ergonomic V-groove structure facilitates measuring coatings on pipes of various sizes.
● Large flat base: Special large flat base probe design ensures stable measurements and higher repeatability.
● Anti-slip pad: Professionally designed anti-slip pad with textured back casing reduces fatigue during prolonged use, reflecting the customer-first philosophy.
III. Probe Parameters Table
1). Probe models: F1.2; N1.2; F3; F5; F10; F0.5
2). Measurement range: F1.2 probe (0-1250μm)
N1.2 probe (0-1250μm)
F3 probe (0-3000μm)
F5 probe (0-5000μm)
F10 probe (0-10000μm)
F0.5 probe (0-500μm)
3). Resolution (μm): 0.1μm
4). Measurement accuracy (H is thickness value):
F0.5 probe/N0.5 probe/F1.2 probe/N1.2 probe accuracy: ±(1~3)% H±2μm
F3 probe/F5 probe accuracy: ±(2~3)% H±2μm
F10 probe accuracy: ±(2~4)% H±2μm
5). Display accuracy: 0~99.9μm: 0.1μm
100~999μm: 1μm
1mm~10mm: 0.01mm
6). Minimum substrate flat diameter: 18mm
7). Minimum curvature radius (concave): 10mm
8). Minimum curvature radius (convex): 5mm
9). Corresponding internal anti-corrosion probe measurable pipe inner diameter: 80mm
10). Probe connection: Split-type cable connection (only replaceable with factory-original range probes).
11).Measurement modes: Simplified mode, monitoring mode, statistical mode, minimum value capture.
12).Operating conditions: Relative humidity: ≤90%; Temperature: -10℃~+40℃.
13).Host dimensions: 150mm(L)*68mm(W)*33mm(H).
14).Power supply: Two AA batteries.
15).Total weight: 220g (including batteries).
IV. MC-3001 Coating Thickness Gauge Standard Packing List
1. MC-3001 Host One Unit
2. AA Batteries Two Pieces
3. Carrying Case One Unit
4. User Manual, Warranty Card, Certificate of Conformity One Set
| Industry Category | Measurement-Analysis-Instruments |
|---|---|
| Product Category | |
| Brand: | 科电 |
| Spec: | |
| Stock: | 10000 |
| Manufacturer: | |
| Origin: | China / Beijing / Chaoyangqu |